Connect Systems CS1000 Spécifications Page 22

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 112
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 21
ContaminationSensor CS 1000 CS1000 hydraulic installation
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 22/112
BeWa CS1000 3764916 300 en-us 2012-08-29.doc 2012-08-29
Selecting the measurement point
In order to obtain cleanliness values that are continuous and coherent in real time,
select a suitable measuring point according to the following guidelines:
A
B
2
A
B
1
3
A
2
1
B
WRONG WRONG OK
1
Select the measurement point so that the sample measured comes from a
turbulent location, with a good flow. For example: on a pipe elbow, etc.
2
Install the sensor near the measurement point to achieve as timely results
as possible.
3
During installation, avoid creating a "siphon" trap for particle deposits in the
line (sedimentation).